Prevention & lipidsCohort study

Method validation of cardiopulmonary exercise testing: intra- and inter-device comparisons across four metabolic carts

Eur Heart J Qual Care Clin OutcomesPublished

  • prospective, single-centre, within-subject, repeated-measures, multi-device comparison study
  • randomized order
  • ramp protocols on cycle ergometer
  • standardized calibration, masks, and timing

Authors' conclusion

Under standardized conditions, the tested metabolic carts yield comparable values both for intra-device reliability and inter-device comparability.

Study context

Population
Healthy adults (35 ± 11 years; 59% males)each completed eight cardiopulmonary exercise testing (CPETs): two per cart, on four systems
Participants
34 adults
Exposure
intra-device reliability and inter-device comparability of CPET parameters

Main finding

As reported in the abstract

Intra-device reliability: VO2, heart rate or workload at anaerobic threshold (AT), RCP or peak
  • Intra-device reliability was high, with no significant differences

Observational data: associations, not effects.

Further findings

OutcomeMGC Ultima™ CPXCOSMED Quark CPET™Schiller PowerCube® ErgoVyaire Vyntus™ CPX
Intra-device: resting VO2Not reported in the abstract56 mL/minNot reported in the abstractNot reported in the abstract
Intra-device: peak ventilation9 L/minNot reported in the abstractNot reported in the abstractNot reported in the abstract
Intra-device: resting VO2 — COSMED resting VO2 variance P
P = 0.0427
Intra-device: peak ventilation — Medical Graphic peak ventilation variance P
P < 0.05

Inter-device: peak VO2 and VE/VCO2 slope up to respiratory compensation point (RCP)

No differences between devices

Inter-device: VE/VCO2 slope (full slope)

Significant differences between machines
P = 0.0048

Inter-device: VO2 at 25%, 50%, 75%, and 100% of ISO-Wpeak

VO2 values largely comparable between devices; significant difference only for VO2 (mL/kg/min) at 25% ISO-Wpeak
P = 0.022

Original publication

Method validation of cardiopulmonary exercise testing: intra- and inter-device comparisons across four metabolic carts. Eur Heart J Qual Care Clin Outcomes. Published 6 March 2026. doi:10.1093/ehjqcco/qcag036

Summarised from the published abstract; every number is checked against it