Coronary disease & ACSRandomised controlled trial

Micro-axial flow pump vs veno-arterial extracorporeal membrane oxygenation for high-risk percutaneous coronary interventions: a randomized trial

Eur Heart JPublished

  • Randomised controlled trial
  • prospective
  • multicentre
  • open-label
  • non-inferiority

Authors' conclusion

The prophylactic use of a micro-axial flow pump during high-risk PCI procedures was non-inferior to VA-ECMO in the incidence of 30-day major adverse events.

Study context

Population
Patients with complex three-vessel disease, unprotected left main disease or last patent conduit and left ventricular ejection fraction (LVEF) ≤ 35% undergoing non-emergent high-risk percutaneous coronary intervention (PCI)
Participants
109 vs 113
Comparison
Micro-axial flow pump vs veno-arterial extracorporeal membrane oxygenation (VA-ECMO) during high-risk PCI
  • SynFlow 3.0n = 109
  • VA-ECMOn = 113reference
Primary endpoint
Major adverse events:
  • All-cause death
  • MI
  • Stroke/TIA
  • Repeat revascularization
  • Major bleeding
  • Acute kidney injury
  • CPR/cardioversion
  • Cardiac surgery/limb ischaemia
  • Serious device-related complications
Follow-up
30 days

Primary result

As reported in the abstract

Absolute difference

−4.6%

95% CI −12.6% to 3.5%

P < 0.001 for non-inferiority

Events by treatment group

SynFlow 3.0 vs VA-ECMO
absolute difference −4.6% (95% CI −12.6% to 3.5%); P for non-inferiority < 0.001

Further findings

Post-procedural hospital stay

  • Post-procedural hospital stay: P = .047

Safety

Device-related adverse events

Anaemia

  • Device-related adverse events: P = .041
  • Anaemia: P = .023

Original publication

Micro-axial flow pump vs veno-arterial extracorporeal membrane oxygenation for high-risk percutaneous coronary interventions: a randomized trial. Eur Heart J. Published 29 July 2026. doi:10.1093/eurheartj/ehag520