Valvular & structural heart diseaseCohort study

Mid-Term Outcomes of a Novel Transcatheter Aortic Valve Replacement (TAVR) Device for Pure Aortic Regurgitation

JACC Cardiovasc IntervPublished

  • Prospective single-arm clinical trial
  • multicentre
  • 15 centres
  • consecutive patients

Authors' conclusion

This study demonstrates that the Pioneer Valve provides a safe and effective alternative for high-risk or inoperable patients with severe pure AR, achieving favorable mid-term outcomes with stable hemodynamic performance.

Study context

Population
Patients with severe pure native aortic regurgitation (AR), high surgical risk or inoperable
Participants
110 patients

Main finding

As reported in the abstract

Mid-term efficacy and safety (VARC-3), up to 1 year · up to 1 year

Technical success

109 of 110 (99.1%)

Valve migration

1 case

Mean aortic valve gradient at 30 days

7.88 ± 2.7 mm Hg

Mean effective orifice area at 30 days

2.2 ± 0.64 cm²

Observational data: associations, not effects.

Further findings

All-cause mortality · up to 1 year

All-cause mortality
3 of 110 (2.7%; 95% CI: 0.9%-8.3%)

Safety

New permanent pacemaker implantation

New permanent pacemaker implantation
23 of 105 cases (21.9%; 95% CI: 14.3%-29.8%)

Original publication

Mid-Term Outcomes of a Novel TAVR Device for Pure Aortic Regurgitation. JACC Cardiovasc Interv. Published 22 December 2025. doi:10.1016/j.jcin.2025.10.017

Summarised from the published abstract; every number is checked against it