Arrhythmias & electrophysiologyCohort study

Multicenter Study on the Safety of Pulsed Field Ablation in Over 40,000 Patients

J Am Coll CardiolPublished

  • Retrospective multicentre registry study
  • 102 US centres

Authors' conclusion

In a real-world setting of unselected U.S. patients, PFA demonstrated a safety profile consistent with preferentiality to functional myocardial tissue ablation, without evidence of esophageal fistula or pulmonary vein stenosis. The major complication rate was ∼0.6%-mostly vascular AEs and pericardial tamponade. Stroke (∼1 in 1,000) and death (∼1 in 2,000) were rare.

Study context

Population
Unselected US patients undergoing AF ablation with the pentaspline pulsed field ablation (PFA) catheter, February 2024–July 2025
Participants
41,968 patients · 102 centres
Outcome
Incidence of major and minor procedure-related AEs

Main finding

As reported in the abstract

Major adverse events (AEs)
0.63%
Cardiac tamponade
0.16%
Vascular injury requiring intervention
0.18%
Stroke
0.10%
Esophageal fistula, persistent phrenic nerve paralysis or pulmonary vein stenosis
0
Mortality at 30 days
0.04%
Unexplained sudden death/cardiac arrest (potential signal)
0.019%
Coronary spasm
0.10%
Acute renal failure requiring dialysis
0.02%
Minor complications
2.05%
Minor vascular issues
0.96%
Pericarditis
0.52%
Self-limited esophageal dysmotility
0.04%

Observational data: associations, not effects.

Further findings

Procedural characteristics

First-time ablation
73%
Paroxysmal AF
54%
Persistent AF
37%
Pulmonary vein isolation
93%
Posterior wall lesions
57%
Cavotricuspid isthmus lesions
31%
Mitral isthmus lesions
14%