Arrhythmias & electrophysiologyCohort study

One-Year Outcomes of the First 1000 Patients Implanted With the Medtronic Micra AV Leadless Pacing System in France: The AV-CESAR Cohort Study

Circ Arrhythm ElectrophysiolPublished

  • nationwide retrospective cohort

Authors' conclusion

The Micra AV system is primarily used as an alternative when dual-chamber conventional pacemakers are not feasible. It demonstrated acceptable safety and clinical performance in a high-risk population. The marked reduction in atrioventricular synchrony at higher heart rates may suggest limited clinical benefit in patients with higher exertional heart rates.

Study context

Population
First patients implanted with a Medtronic Micra AV leadless pacemaker in France (2020–2024)device implantation successful in 1000 (99.7%)
Participants
1,003 patients
Outcome
Early (in-hospital) and late (postdischarge) device-related major complications
Follow-up
Mean 13.4±10.6 months

Main finding

As reported in the abstract

Early major complications
2.4%
Early: pericardial effusion (3 requiring surgery)
0.8%
Early: access-site complications
0.8%
Early: thromboembolism
0.3%
Late: pacemaker syndrome
2.5%
Late: pacing-induced cardiomyopathy
1.2%
Late: high thresholds (>4 V/0.24ms)
0.3%
Late: premature battery depletion
0.2%

Observational data: associations, not effects.

Further findings

Implantation of a new pacing system

Required implantation of a new pacing system
2.5%
pacing-induced cardiomyopathy
n=11
pacemaker syndrome
n=7
high threshold
n=3
battery/software failure
n=3
tricuspid regurgitation
n=1
Permanently reprogrammed to VVI mode
12.7%

Atrioventricular synchrony (pacing-dependent patients with ambulatory Holter monitoring, >3.4 million paced cycles)

Mean atrioventricular synchrony
67.6±17.3%
Device-reported AM-VP
74.8±17.1%
Correlation R 2
0.92
P
P < 0.001
Atrioventricular synchrony declined at heart rates >90 beats per minute (share of monitored time)
5.6%

All-cause mortality

All-cause mortality
13.8%
Directly device-related
0.2%
Deaths that were cardiovascular
41.7%

Baseline characteristics

Mean age
72.0±15 years
Men
62.8%
≥2 comorbidities
44.0%
Unsuitable for transvenous pacemakers
62.4%
Primary indication: permanent complete atrioventricular block with preserved sinus rhythm
78.0%

Original publication

One-Year Outcomes of the First 1000 Patients Implanted With the Medtronic Micra AV Leadless Pacing System in France: The AV-CESAR Cohort Study. Circ Arrhythm Electrophysiol. Published 20 May 2026. doi:10.1161/circep.126.014850

Summarised from the published abstract; every number is checked against it