Optimization of manual ventilation quality using respiratory function monitoring in neonates: A two-phase intervention trial
Main finding · pVTeResult — go to the resultMean difference 16.95 percentage points CI 7.4–35Authors' conclusion
Our results suggest that the clinical application of a RFM for manual ventilation of preterm and term infants leads to a significant improvement in ventilation parameters.
Study context
- Population
- Neonates (GA 22–66 weeks) needing positive pressure ventilation, NICU and delivery room, Medical University of Vienna
- Participants
- 90 neonates
- Comparison
- Visible respiratory function monitor (RFM) vs hidden RFM during manual positive pressure ventilation
- Outcome
- Percentage of ventilations within a tidal volume range of 4-8 ml/kg (pVTe)
Methodological detailssingle-centre, two consecutive phases, Each phase approximately six months
- Single-centre
- Two consecutive phases
- Each phase approximately six months
Main finding
pVTe
Interventionvs Control
mean difference 16.95 percentage points (CI 7.4–35); P < 0.001
As reported in the abstract
Key notes
- Printed as "16.95% CI: 7.4-35"
Secondary outcomes
| Outcome | Intervention | Control | P value |
|---|---|---|---|
| Excessive tidal volumes (>8 ml/kg) | 10.9% (IQR 26.4) | 29.5% (IQR 38.1) | P = 0.004 |
- Mask leakage
- P = 0.002