Optimization of manual ventilation quality using respiratory function monitoring in neonates: A two-phase intervention trial

ResuscitationPublished Two-phase intervention study

Main finding · pVTeResult — go to the resultMean difference 16.95 percentage points CI 7.4–35

Authors' conclusion

Our results suggest that the clinical application of a RFM for manual ventilation of preterm and term infants leads to a significant improvement in ventilation parameters.

Study context

Population
Neonates (GA 22–66 weeks) needing positive pressure ventilation, NICU and delivery room, Medical University of Vienna
Participants
90 neonates
Comparison
Visible respiratory function monitor (RFM) vs hidden RFM during manual positive pressure ventilation
Outcome
Percentage of ventilations within a tidal volume range of 4-8 ml/kg (pVTe)
Methodological detailssingle-centre, two consecutive phases, Each phase approximately six months
  • Single-centre
  • Two consecutive phases
  • Each phase approximately six months

Main finding

pVTe

Mean difference

16.95 percentage points

CI 7.4–35

P < 0.001

By group

Interventionvs Control
mean difference 16.95 percentage points (CI 7.4–35); P < 0.001

As reported in the abstract

Secondary outcomes

OutcomeInterventionControlP value
Excessive tidal volumes (>8 ml/kg)10.9% (IQR 26.4)29.5% (IQR 38.1)P = 0.004
Mask leakage
P = 0.002

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Optimization of manual ventilation quality using respiratory function monitoring in neonates: A two-phase intervention trial. Resuscitation. Published 2 August 2024. doi:10.1016/j.resuscitation.2024.110345

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.