Arrhythmias & electrophysiology

Optimizing Pulsed Field Ablation

JACC Clin ElectrophysiolPublished

  • single-center study
  • consecutive atrial fibrillation (AF) patients
+1 more
  • consecutive atrial fibrillation (AF) patients

Authors' conclusion

In this large cohort, PFA demonstrated high efficacy with short procedure times, low complication rates, and favorable long-term outcomes. Workflow refinements improved safety and PVI durability, underscoring PFA's role for standardized and safe AF ablation.

Study context

Population
Consecutive AF patients undergoing pulmonary vein isolation (PVI) using the pentaspline pulsed field ablation (PFA) catheter between April 2021 and May 2024, excluding ablation beyond PVIpatients included
Participants
1,000 patients
Comparison
Sequential workflow refinements, including the transition from 8 applications to the 10-application "olive" strategy

Main finding

As reported in the abstract

PVI achieved

median procedure time

30 minutes (Q1-Q3: 25-40 minutes)

Further findings

Outcomeparoxysmal AFpersistent AFconventional approacholive strategyP value
Arrhythmia-free survival · One-year · higher in paroxysmal AF83.8%66.6%Not reported in the abstractNot reported in the abstractP < 0.001conventional approach, olive strategy: not reported in the abstract
PVI durability by approach · increased with the olive strategyNot reported in the abstractNot reported in the abstract78.7%91.6%P < 0.001paroxysmal AF, persistent AF: not reported in the abstract
Arrhythmia-free survival — overall
76.6%
  • Arrhythmia-free survival: Median follow-up time was 454 days (Q1-Q3: 370-631 days)

Durable PVI at repeat procedures

veins with durable PVI across 173 repeat procedures
82% (n = 562 of 685)

PVI durability by approach

PVI durability increased from 78.7% with the conventional approach to 91.6% with the olive strategy, while long-term outcomes remained unchanged

Safety

Complications

major complication rate
0.5%
total complication rate (primarily access site related)
2.1%

Tamponade, before vs after workflow refinements

before refinements
0.8% (n = 2 of 260)
after refinements
0% (n = 0 of 740)
  • trend toward reduction (P = 0.067)

Procedure-related bradycardia, before vs after workflow refinements

before refinements
49.3%
after refinements
5%
  • P < 0.001

Original publication

Optimizing Pulsed Field Ablation. JACC Clin Electrophysiol. Published 22 July 2026. doi:10.1016/j.jacep.2026.06.011

Summarised from the published abstract; every number is checked against it