Coronary disease & ACS

Paclitaxel-Coated Balloon for the Treatment of Small Vessel In-Stent Restenosis: A Subgroup Analysis of the AGENT IDE Randomized Trial

JACC Cardiovasc IntervPublished Secondary analysis of the AGENT IDE randomised trial

Authors' conclusion

This prespecified subgroup analysis demonstrates that angioplasty with a PCB was associated with consistently reduced rates of 1-year TLF compared with an uncoated balloon in both SV and large vessel ISR patients.

Study context

Population
Patients with coronary in-stent restenosis (ISR), small vessel (RVD ≤2.75 mm) or large vessel (RVD >2.75 mm)597 with known vessel size
Participants
600 patients
Comparison
Paclitaxel-coated balloon vs uncoated balloon angioplasty, by vessel size
Primary endpoint
Target lesion failure (TLF) (target lesion revascularization, cardiac death, target vessel-related MI)
Follow-up
1 year
Methodological detailsprespecified subgroup analysis, 2:1 randomisation
  • Prespecified subgroup analysis
  • 2:1 randomisation

Primary result

As reported in the abstract

1-year TLF by vessel size
PCB vs Uncoated balloonHazard ratio (95% CI)
Small vessels: 17.7%vs 27.4%; 39% relative reduction
0.61(0.37–0.99)
Large vessels: 18.4%vs 30.5%; 43% reduction
0.57(0.34–0.96)

Further findings

OutcomeSmall vesselsLarge vesselsHR (95% CI)P value
1-year TLF, smallvs large vessels (all patients)20.6%22.6%0.92 (0.65–1.31)P = 0.65

Safety

Definite or probable stent thrombosis — PCB
0 events

Original publication

Paclitaxel-Coated Balloon for the Treatment of Small Vessel In-Stent Restenosis: A Subgroup Analysis of the AGENT IDE Randomized Trial. JACC Cardiovasc Interv. Published 24 November 2025. doi:10.1016/j.jcin.2025.09.032

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.