Heart failure & cardiomyopathiesCohort study

Physician-Directed Patient Self-Management Using an Implantable Inferior Vena Cava (IVC) Congestion Sensor

JACC Heart FailPublished

  • Prospective cohort study
  • 2 sequential prospective cohorts
  • early feasibility study

Authors' conclusion

Physician-directed patient self-management using ambulatory IVC-based congestion monitoring was feasible, associated with increased signal-driven intervention, and not associated with adverse safety signals. This strategy demonstrated favorable clinical signals.

Study context

Population
Heart failure (HF) patients with an implantable IVC-based congestion management system (FUTURE-HFII)
Participants
10 vs 15
Exposure
Self-management vs Clinician-managed care
  • Self-managementn = 10
  • Clinician-managed caren = 15reference
Outcome
Safety, device adherence, medication adjustments, congestion metrics, HF events and quality of life
Follow-up
90 days

Main finding

As reported in the abstract

  • No safety signals were observed

Observational data: associations, not effects.

Further findings

OutcomeRRP value
Diuretic changes per patient (signal intervention frequency) · 10.4 vs 2.8 diuretic changes per patient3.71P < 0.01

Daily device usage

High in both groups, numerically greater with self-management
90.5% vs 84.8%
  • 1,927 measurements

Congestion control (NORM score time-in-range)

Comparable between groups

Quality of life (KCCQ-CSS gain ≥5 points)

Improved in both groups, larger gains in the self-managed cohort
60.0% vs 46.7%

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Physician-Directed Patient Self-Management Using an Implantable IVC Congestion Sensor. JACC Heart Fail. Published 3 March 2026. doi:10.1016/j.jchf.2026.103011

Summarised from the published abstract; every number is checked against it