Predicting worsening heart failure hospitalizations in patients with implantable cardioverter defibrillators: is it all about alerts? A pooled analysis of nine trials

EuropacePublished Pooled analysis of nine clinical trials

Main finding · Mean HF scoreResult — go to the resultDifference +22%

Authors' conclusion

The HF score was significantly higher at, and further increased during 12 weeks before WHFH, as compared to the no WHFH group, with seven components showing different behaviour and contribution.

Study context

Population
Patients with a defibrillator capable of atrial sensing and remote monitoring, EF ≤ 35%, NYHA class II/III, no long-standing AF
Participants
2,050 patients9 trials
Comparison
WHFH vs no WHFH
Outcome
Temporal trends of the HeartInsight HF score and its seven components before WHFH vs 12-week trends without WHFH
Follow-up
12 weeks before WHFH

Main finding

Mean HF score

Difference

+22%

P = 0.003

By group

WHFH

42.3

mean · SD 26.1

No WHFH

30.7

mean · SD 20.6

As reported in the abstract

Further findings

Algorithm components vs no WHFH

Already higher 12 weeks before WHFH: 24 h HR, HR variability, thoracic impedance
Significantly increased until WHFH: nocturnal HR, atrial tachyarrhythmia, ventricular extrasystoles, patient activity

Original publication

Predicting worsening heart failure hospitalizations in patients with implantable cardioverter defibrillators: is it all about alerts? A pooled analysis of nine trials. Europace. Published 31 January 2024. doi:10.1093/europace/euae032

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.