Arrhythmias & electrophysiologyResearch article

Pulsed-Field Ablation to Enhance the Durability of Pulmonary Vein Isolation and Treat Paroxysmal Atrial Fibrillation

J Am Coll CardiolPublished

  • Single-arm trial
  • multicentre
  • pivotal investigational device exemption trial

Authors' conclusion

The favorable clinical success observed in this multicenter Food and Drug Administration trial of this novel PFA catheter appears to reflect the clinical success achievable with durable PVI in PAF patients.

Study context

Population
Patients with symptomatic drug-resistant or intolerant paroxysmal AF19 roll-in, 164 pivotal
Participants
183 patients
Outcome
Freedom from atrial arrhythmias and from other failure modes (failure to achieve PVI, repeat ablation or cardioversion, new or increased class I/III antiarrhythmic drugs)
Follow-up
12 months

Main finding

As reported in the abstract

Primary effectiveness (freedom from atrial arrhythmias and other failure modes)

Primary effectiveness at 12 months

77.8% (95% CI: 70.6%-83.9%)

Further findings

Pulmonary veins (PVs) isolated by first pulsed field application

PVs electrically isolated by first application
94%, 607/647 veins

Durable PV isolation at redo ablation (11 patients)

PVs durably isolated at redo ablation
39 of 41 (95.1%)

Procedure times

Mapping time, min
9.0 ± 5.5
Transpired ablation time, min
25.6 ± 9.8
Left atrial catheter dwell time, min
59.9 ± 11.6
Procedure time, min
95.8 ± 20.7

Safety

Primary safety events within 7 days

Primary safety event rate (hemorrhagic stroke)
0.6% (1/164)

PV stenosis (substudy)

Moderate or severe PV narrowing
0

Original publication

Pulsed-Field Ablation to Enhance the Durability of Pulmonary Vein Isolation and Treat Paroxysmal Atrial Fibrillation. J Am Coll Cardiol. Published 21 January 2026. doi:10.1016/j.jacc.2025.10.084

Summarised from the published abstract; every number is checked against it