Evaluation of the R-One robotic system for percutaneous coronary intervention: the R-EVOLUTION study
Authors' conclusion
This study suggests that robotic PCI using R-One is safe and effective with markedly lower radiation exposure to the operator.
Study context
- Population
- Patients with a de novo coronary artery stenosis (length <38 mm, reference diameter 2.5–4.0 mm) undergoing stent implantation; recent STEMI, ostial or left main lesion, bifurcation, severe tortuosity or calcification excluded
- Participants
- 62 patients64 lesions
- Outcome
- Clinical success (absence of major intraprocedural complications) and technical success (successful advancement and retraction of all PCI devices without total manual conversion)
Methodological detailsmulticentre, European
- Multicentre
- European
Main finding
As reported in the abstract
Clinical success
100%
Technical success
95.2%
59/62
Observational data: associations, not effects.
Further findings
Radial access
- Radial access
- 96.8% (60/62)
Robotic procedure duration
- Mean robotic procedure duration
- 19.9±9.6 min
Fluoroscopy time
- Mean fluoroscopy time
- 10.3±5.4 min
Total manual conversion
- Total manual conversion
- 4.8% (3/62)
- Directly related to the robotic system
- 1
Operator radiation exposure vs doses on the patient table
- Reduction under the lead apron
- 84.5%
- Reduction on top of the lead apron
- 77.1%
Safety
- No complications at 30 days