Heart failure & cardiomyopathiesCohort study

Radiofrequency current–facilitated implantation of left bundle branch pacing lead for cardiac resynchronization therapy

Heart RhythmPublished

  • Case series

Authors' conclusion

RF-facilitated LBBP lead implantation is a feasible technique with potential cautious use as a bailout in selected cases, especially for failed CRT delivery.

Study context

Population
Patients with challenging left bundle branch pacing (LBBP) lead implantation, most cardiac resynchronization therapy (CRT) eligible, in whom radiofrequency (RF) current was delivered to the lead tip for septal electrosection37 RF-facilitated lead implantations in 36 patients
Participants
36 patients

Main finding

As reported in the abstract

RF use per implantation

RF applications per implantation

2.1 ± 1.7

Seconds of RF

2.6 ± 2.2 seconds

Median power

27 W

Observational data: associations, not effects.

Further findings

RF delivery used to achieve a current of injury drop not obtainable with manual lead rotation alone

Share of cases
85.7%

Lead implantation success

Successful in all but 2 cases

Lead performance (pacing parameters and lead stability) · follow-up

Follow-up threshold
0.8 ± 0.3 V
Sensitivity
12.0 ± 5.6 mV
  • good and did not seem to be affected by RF use

Follow-up duration

Mean follow-up duration
236 ± 165 days

Safety

Proarrhythmia and complications

Patients with sustained ventricular tachycardia
1
In all other cases proarrhythmia absent or self-limited; no other complications

Original publication

Radiofrequency current–facilitated implantation of left bundle branch pacing lead for cardiac resynchronization therapy. Heart Rhythm. Published 10 April 2026. doi:10.1016/j.hrthm.2026.03.1943

Summarised from the published abstract; every number is checked against it