Arrhythmias & electrophysiologyCohort study

Rapid early battery depletion and lead–electrode corrosion owing to CIED gate oxide semiconductor defects

Heart RhythmPublished

  • Analysis of manufacturers' reports in the Food and Drug Administration's publicly available Manufacturer and User Facility Device Experience database

Authors' conclusion

GO defects represent an early, manufacturing-related semiconductor failure mechanism in CIEDs that can rapidly deplete batteries, disable lifesaving therapy, and contribute to lead-electrode corrosion. Early recognition of characteristic device and clinical patterns may facilitate risk mitigation, particularly in vulnerable patients.

Study context

Population
Cardiac implantable electronic devices (CIEDs) with gate oxide (GO) failures reported by Abbott, Biotronik, Boston Scientific, and MedtronicGO failure reports · Boston Scientific devices (2011-2025) 178 · Medtronic devices (2009-2021) 30
Exposure
Pacemakers vs implantable cardioverter-defibrillators

Main finding

As reported in the abstract

GO failures reported for Abbott or Biotronik
  • none were reported for Abbott or Biotronik

Observational data: associations, not effects.

Further findings

Time to failure

Time to failure — Median time to failure, all
5.8 months (interquartile range [IQR] 2.9-11.6)
  • Time to failure: Pacemakers failed earlier than implantable cardioverter-defibrillators

Presenting findings

premature battery depletion
n = 83
safety mode activation
n = 45
low lead impedance
n = 57
loss of telemetry
n = 25
high power consumption
n = 21

Engineering analyses of explanted leads

lead-electrode corrosion in explanted leads
10
electrical discontinuity, of the 10 corroded leads
4

Safety

Major adverse events

death
n = 2
cardiac arrest/asystole/shock
n = 5
syncope
n = 7
heart failure
n = 6

Original publication

Rapid early battery depletion and lead–electrode corrosion owing to CIED gate oxide semiconductor defects. Heart Rhythm. Published 26 February 2026. doi:10.1016/j.hrthm.2026.02.037

Summarised from the published abstract; every number is checked against it