Risk of occurrence of electromagnetic interference from the application of transcutaneous electrical nerve stimulation on the sensing function of implantable defibrillators

EuropacePublished

  • Pilot safety study
  • standardized test protocol

Authors' conclusion

Overall, only intermittent and minor EMI were detected.

Study context

Population
Patients with modern implantable cardioverter-defibrillator (ICD)/cardiac resynchronization therapy (CRT) defibrillators (55 models, 4 manufacturers) undergoing transcutaneous electrical nerve stimulation (TENS) at cervical spine and thorax, high-frequency (80 Hz) and burst-mode (2 Hz)
Participants
107 patients
Outcome
Occurrence of EMI from TENS on ICD sensing

Main finding

As reported in the abstract

EMI detected
Electromagnetic interference (EMI) detected
17 of 107 patients (15.9%)
Interpretation as premature ventricular beats (VS/S)
15 patients (14%)
Noise reversion
5 (4.6%)
Temporary asynchronous pacing (from noise reversion)
3 (2.8%)
Interpretation as VT/VF
2 (1.9%)
Premature atrial beat
2 (1.9%)

Further findings

Factors influencing EMI occurrence

Position (chest)
P < 0.01
Higher current intensity
P < 0.01
Manufacturer
P = 0.012