Arrhythmias & electrophysiology

Role of the Bridge balloon in improving the safety of transvenous lead extraction procedures

Heart RhythmPublished Retrospective analysis of a prospectively maintained single-centre registry

Authors' conclusion

Introduction of the Bridge balloon was associated with reduced TLE mortality, particularly in electrophysiology laboratory procedures, despite more complex cases.

Study context

Population
Patients undergoing transvenous lead extraction (TLE) at one centre, 1996–2022
Participants
4,946 patients
Exposure
Bridge balloon era vs pre-balloon era
  • Bridge balloon eran = 1,369
  • Pre-balloon eran = 3,577reference
Outcome
Periprocedural mortality
Methodological detailsbefore–after comparison
  • Before–after comparison

Main finding

As reported in the abstract

Procedure-related deaths

Bridge balloon era

0

events

Pre-balloon era

0.3%

12 of 3577

P = 0.03

Observational data: associations, not effects.

Further findings

Intraprocedural death, EP laboratory procedures (subgroup)

P = 0.007

7-day mortality, EP laboratory procedures (subgroup)

P = 0.005

Predictors of prophylactic balloon placement

ORP value
Older lead age1.12P < 0.001
Higher combined age of leads1.04P = 0.003
Defibrillator leads1.50P < 0.001
OutcomeBridge balloon eraPre-balloon era
Bridge balloon use (balloon era)
Prophylactic placement794 (58%)— not stated in the abstract
Rescue activation21 (1.5%)— not stated in the abstract

— not stated in the abstract

Safety

Rescue sternotomy · balloon era had older leads

P < 0.001

Original publication

Role of the Bridge balloon in improving the safety of transvenous lead extraction procedures. Heart Rhythm. Published 26 June 2025. doi:10.1016/j.hrthm.2025.06.038

Composed by AI from the published abstract and checked against it. How Briefs are made →