Congenital heart diseaseCohort study

Routine application of three-dimensional transesophageal echocardiography-based device selection for transcatheter closure of atrial septal defect without balloon sizing

Eur Heart J Cardiovasc ImagingPublished

  • prospective atrial septal defect (ASD)_3D_Device_Size registry
  • consecutive patients

Authors' conclusion

Pre-procedural determination of device size based on 3D TEE provides a safe and useful framework in routine clinical practice. Because of the diversity in maximal diameter orientation, balloon sizing using a fixed fluoroscopic angle may underestimate the true maximal diameter of an ASD. (ClinicalTrials.gov number NCT 02097758)

Study context

Population
Consecutive patients undergoing transcatheter ASD closure with device size determined prior to the procedure by a 3D transesophageal echocardiography (TEE)-based formulapatients included between September 2016 and May 2024
Participants
748 patients

Main finding

As reported in the abstract

Procedural and fluoroscopy times

Median procedural time

18.0 min

Median fluoroscopy time

3.7 min

Observational data: associations, not effects.

Further findings

Procedure success rate

Procedure success rate
99.5%

Intraprocedural device size reselection

Patients requiring intraprocedural device size reselection
One patient (0.1%)

Follow-up

Median clinical follow-up
1.6 years

Oval ASDs and long axis orientation

Oval ASDs
186 (24.9%)
Long axis orientations were diverse

Computational simulation of balloon sizing in fluoroscopic antero-posterior projection

Greatest underestimation of maximal diameter, in horizontally oriented oval ASDs
34.9%

Safety

Device embolization

Device embolization
three (0.4%) patients
  • two of these cases were attributed to delivery system malfunction

Device malposition

Device malposition, underwent surgery
one (0.1%) patient

Cardiac erosion requiring surgery · During follow-up

Patients requiring surgery for cardiac erosion
one patient

Original publication

Routine application of three-dimensional transesophageal echocardiography-based device selection for transcatheter closure of atrial septal defect without balloon sizing. Eur Heart J Cardiovasc Imaging. Published 13 January 2026. doi:10.1093/ehjci/jeag007

Summarised from the published abstract; every number is checked against it