Arrhythmias & electrophysiologyCohort study

Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study

EuropacePublished

  • Prospective single-arm study
  • global
  • multicentre

Authors' conclusion

In this trial, ablation with TactiFlex Duo in patients with symptomatic, drug-refractory PAF was safe and effective.

Study context

Population
Symptomatic, drug-refractory paroxysmal atrial fibrillation (PAF)
Participants
180 participants
Outcome
Safety: pre-defined safety events within 7 days of any ablation procedure; effectiveness: freedom from recurrent AF/atrial tachycardia/atrial flutter, repeat procedure, new or escalated class I/III AAD and/or cardioversion
Follow-up
12 months

Main finding

As reported in the abstract

Freedom from composite of:
  • Recurrent AF/AT/atrial flutter
  • Repeat procedure
  • New or escalated class I/III AAD and/or cardioversion

Primary effectiveness endpoint

74.6%

Observational data: associations, not effects.

Further findings

Freedom from documented AF/AFL/AT recurrence · 12 months

Freedom from documented AF/AFL/AT recurrence
77.4%
Without weekly transtelephonic monitoring
92.6%

Successful pulmonary vein isolation (PVI)

Successful PVI achieved in all participants

Additional non-PV ablations

Subjects with additional non-PV ablations
37.2% (67/180)

Durable PV isolation at remapping (15 subjects with repeat ablation after blanking period)

PVs durably isolated from index procedures
79.7% (47/59)

Freedom from AF/AFL/AT with PFA Index ≥30 and AutoMark distance up to 6 mm target workflows · 12 months

Freedom from AF/AFL/AT
84.6%
  • retrospective analysis

Safety

Pre-defined safety events within 7 days of any ablation procedure (primary safety endpoint)

Primary safety endpoint event rate
1.7%

Original publication

Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study. Europace. Published 29 August 2026. doi:10.1093/europace/euag238

Summarised from the published abstract; every number is checked against it