Arrhythmias & electrophysiologyCohort study

Safety and effectiveness of the TactiFlex duo system: 6-month results of the FOCALFLEX study

EuropacePublished

  • Prospective single-arm study
  • multicentre
  • 21 sites worldwide
  • Enrolment August 2024 – January 2025

Authors' conclusion

The FOCALFLEX study results demonstrate safety and effectiveness of a novel dual-mode RF/PFA system through 6 months of follow-up to treat symptomatic, recurrent PAF.

Study context

Population
Patients with paroxysmal AF
Participants
144 patients · 21 sites
Outcome
Safety: pre-defined serious adverse events within 7 days; effectiveness: freedom from documented AF/AFL/AT episodes >30 s after a 90-day blanking period
Follow-up
7 days (safety); 6 months (effectiveness)

Main finding

As reported in the abstract

Primary effectiveness (freedom from documented AF/AFL/AT >30 s after 90-day blanking) · 6 months

Protocol-defined primary effectiveness rate

79.6%

Observational data: associations, not effects.

Further findings

Freedom from documented arrhythmia recurrence · 6 months

Free from documented arrhythmia recurrence
81.0%

Acute effectiveness (entrance block + post-ablation left atrium (LA) voltage map after 20-min wait)

Treated PVs achieved
99.6% (566/568)
Pulsed field ablation (PFA) applications/patient, mean ± SD
95.5 ± 37.1

Procedure, fluoroscopy, LA dwell and PV ablation times

Procedure time
112.2 ± 29.5 min
Fluoroscopy time
7.6 ± 5.7 min
LA dwell time
81.1 ± 19.7 min
PV ablation time
47.4 ± 16.4 min

Repeat ablation after blanking period · through 6 months

Repeat ablation
1 (0.7%)

Safety

Primary safety events (pre-defined serious AEs within 7 days)

Primary safety events
2 (1.4%)

Original publication

Safety and effectiveness of the TactiFlex duo system: 6-month results of the FOCALFLEX study. Europace. Published 7 May 2026. doi:10.1093/europace/euag108

Summarised from the published abstract; every number is checked against it