Safety and effectiveness of transvenous extraction for ePTFE-coated leads: A retrospective comparative study

Heart RhythmPublished

  • single-center, retrospective study
  • matched control cohort sensitivity analysis (Mahalanobis distance nearest-neighbor matching)
+1 more
  • matched control cohort sensitivity analysis (Mahalanobis distance nearest-neighbor matching)

Authors' conclusion

The safety and efficacy of extraction of Reliance ICD leads were similar to the extraction of non-Reliance ICD leads. Reliance procedure times were shorter, and there was less rotational cutting tool use.

Study context

Population
Patients undergoing extraction of Reliance and non-Reliance implantable cardioverter-defibrillator (ICD) leadsReliance patients (88 leads) 87 · non-Reliance patients (controls, 257 leads) 220
Exposure
extraction of Reliance and non-Reliance implantable cardioverter-defibrillator (ICD) leads
  • Reliancen = 87 patients (88 leads)
  • non-Reliance (controls)n = 220 patients (257 leads)
Outcome
Extraction failure rate

Main finding

As reported in the abstract

Observational data: associations, not effects.

Further findings

Mean, years

OutcomeReliancenon-Reliance (controls)matched controls
Age58.5 (SD 14.4)60.1 (SD 14.5)Not reported in the abstractmatched controls: not reported in the abstract
  • Age: P = .33

Body mass index, diabetes mellitus, left ventricular ejection fraction, previous sternotomy, estimated glomerular filtration rate

Similar in both groups
P > 0.05 for all

Procedural characteristics (adjusted)

Reliance procedures required less extraction time, less rotational cutting tool use, and fewer clicks with similar use and amount of laser
  • After adjustment

Predictors of abnormal impedance among Reliance leads

High previous impedance, dual-coil configuration, and low body mass index were predictors of abnormal impedance

Safety

Major adverse event rate (main safety endpoint)

  • Major adverse event rate (main safety endpoint): Reliance vs control P = .72
  • Major adverse event rate (main safety endpoint): Reliance vs matched controls P = .62

Major adverse event rate (main safety endpoint)

No difference in the major adverse event rate

Original publication

Safety and effectiveness of transvenous extraction for ePTFE-coated leads: A retrospective comparative study. Heart Rhythm. Published 30 May 2026. doi:10.1016/j.hrthm.2026.05.043

Summarised from the published abstract; every number is checked against it