Safety and effectiveness of the first balloon-in-basket pulsed field ablation system for the treatment of atrial fibrillation: VOLT CE Mark Study 6-month results

EuropacePublished Prospective single-arm study

Authors' conclusion

The VOLT CE Mark Study primary results demonstrate the safety and effectiveness of the novel balloon-in-basket PFA system to perform PVI in PAF and PersAF.

Study context

Population
Patients with drug-refractory paroxysmal or persistent atrial fibrillation (AF) undergoing pulmonary vein isolation (PVI) with a balloon-in-basket pulsed field ablation (PFA) system146 underwent PVI
Participants
150 patients
Exposure
PAF vs persistent AF (PersAF)
Outcome
Freedom from documented atrial arrhythmias
Follow-up
6 months
Methodological detailsmulticentre, pre-market
  • Multicentre
  • Pre-market

Main finding

As reported in the abstract

Freedom from documented atrial arrhythmias · 6 months

Freedom from symptomatic recurrence

Acute effectiveness, treated PVs
99.1% (573/578)
Acute effectiveness, patients
98.6% (144/146)

Observational data: associations, not effects.

Further findings

Procedural metrics

PFA applications/patient
17.6 ± 5.7
Procedure time (min)
100.4 ± 33.0
Fluoroscopy time (min)
17.3 ± 12.1
LA dwell time (min)
39.4 ± 20.6
Transpired ablation time (min)
31.4 ± 16.8

Safety

Primary serious adverse events · within 7 days

Primary serious adverse events
4 (2.7%; 4/146)

Original publication

Safety and effectiveness of the first balloon-in-basket pulsed field ablation system for the treatment of atrial fibrillation: VOLT CE Mark Study 6-month results. Europace. Published 31 March 2025. doi:10.1093/europace/euaf072

Summarised from the published abstract; every number is checked against it