Safety, efficacy, and functional outcomes using a novel small-diameter defibrillation lead for left bundle branch area pacing: Final results of the LEADR left bundle branch area pacing (LBBAP) clinical trial

Heart RhythmPublished

  • clinical trial

Authors' conclusion

The final study results of the lead evaluation for defibrillation and reliability in left bundle branch area pacing (LEADR LBBAP) clinical trial demonstrate continued stable electrical performance, a low rate of lumenless defibrillation lead-related major complications, and the benefit of LOT-CRT when a lumenless defibrillation lead is implanted for LBBAP.

Study context

Population
Patients with a de novo indication for an implantable cardioverter-defibrillator or CRT-D undergoing LBBAP with the OmniaSecure defibrillation leadpatients with an LBBAP implantation attempt · LBBAP-optimized CRT (LOT-CRT) 143
Participants
316 patients

Main finding

As reported in the abstract

Electrical performance (pacing capture threshold, impedance, R-wave amplitude) · follow-up 9.9 ± 4.0 months
  • Remained stable

Further findings

Freedom from study lead-related major complications · 6 and 12 months

6 months
97.7%
12 months
96.8%

Appropriate therapy (shock or antitachycardia pacing) · 1 year

Patients receiving appropriate therapy
7.9%

Inappropriate shock rate · 1 year

Inappropriate shock rate
2.1%

Paced QRS duration (QRSd) vs intrinsic QRSd (LOT-CRT patients) · 6 months

Significant reduction in paced QRSd compared with intrinsic QRSd
P < 0.001

Left ventricular ejection fraction (LOT-CRT patients) · baseline to 6 months

Improved
P < 0.001

Indexed left ventricular end-systolic volume (LOT-CRT patients) · baseline to 6 months

Reduced
P < 0.001

Quality of life (LOT-CRT patients) · baseline to 6 months

Kansas City Cardiomyopathy Questionnaire improved
P < 0.001
Clinical composite score: improved or stabilized
85%

Original publication

Safety, efficacy, and functional outcomes using a novel small-diameter defibrillation lead for left bundle branch area pacing: Final results of the LEADR LBBAP clinical trial. Heart Rhythm. Published 26 August 2026. doi:10.1016/j.hrthm.2026.08.030

Summarised from the published abstract; every number is checked against it