Arrhythmias & electrophysiology

Safety of ventricular arrhythmia radiofrequency ablation with half-normal saline irrigation

EuropacePublished Prospective cohort study

Authors' conclusion

Half-normal saline irrigation RF ablation with power guided by impedance fall and intracardiac echocardiography has an acceptable rate of complications and acute ablation success while administering half of the saline load expected for NS irrigation.

Study context

Population
Patients undergoing endocardial and/or epicardial radiofrequency (RF) ablation of ventricular arrhythmiasprocedures in 935 patients
Participants
1,024 patients
Exposure
Half-normal saline vs normal saline irrigation for RF ablation
  • Half-normal saline (HNS)n = 900
  • Normal saline (NS)n = 124reference
Outcome
Periprocedural complications and acute ablation outcome
Follow-up
30 days
Methodological detailsconsecutive patients, Irrigant selected at the discretion of the treating physician, RF power titrated to a ≤15 Ω impedance fall with intracardiac echocardiography monitoring
  • Consecutive patients
  • Irrigant selected at the discretion of the treating physician
  • RF power titrated to a ≤15 Ω impedance fall with intracardiac echocardiography monitoring

Main finding

As reported in the abstract

Any adverse event

HNS

13.0%

Observational data: associations, not effects.

Safety

Stroke/TIA (HNS) · 30 days
0.4% · 4 of 900
Pericardial effusion requiring treatment (HNS) · 30 days
3.8% · 34 of 900

Steam pops with perforation requiring surgical repair

Steam pops with perforation requiring surgical repair
2 (0.2%)

Original publication

Safety of ventricular arrhythmia radiofrequency ablation with half-normal saline irrigation. Europace. Published 17 February 2024. doi:10.1093/europace/euae018

Composed by AI from the published abstract and checked against it. How Briefs are made →