Arrhythmias & electrophysiologySecondary analysis

Sensing and detection performance of the novel, small-diameter OmniaSecure defibrillation lead: in-depth analysis from the LEADR trial

EuropacePublished

  • Secondary analysis of the LEADR trial
  • single-group
  • longer-term follow-up
  • Mean follow-up 18.2 ± 5.5 months

Authors' conclusion

Chronic sensing performance of the OmniaSecure defibrillation lead demonstrated R-wave stability with a low rate of P-wave and T-wave oversensing, resolved predominantly by adjusting RV sensitivity. Further, VT/VF detection was successful and was not impacted when programmed to less sensitive settings.

Study context

Population
Patients with indications for de novo ICD/CRT-D643 (97.9%) successfully implanted
Participants
657 patients
Primary endpoint
Sensing and detection performance of the OmniaSecure lead

Primary result

As reported in the abstract

Sensing and detection performance
PWOS-related adverse event, none with inappropriate shock
4 patients (0.6%)
TWOS-related adverse event
4 patients (0.6%)
TWOS with inappropriate shock
3 patients (0.5%)
Appropriate VF detection at implant at least sensitive setting of 1.2 mV
97.6% (120/123)
VT/VF episodes appropriately detected and treated in follow-up, no under-detected episodes
670 episodes in 94 patients

Further findings

Freedom from major study lead-related complications · 24 months

Freedom from major study lead-related complications at 24 months
96.9%

Appropriate therapies (shock and/or ATP) · 24 months

Patients receiving appropriate therapies (shock and/or ATP) at 24 months
17.6%
ATP efficacy
76.5%

Lead fractures

Fractures during follow-up
0

VF detection at implant (least sensitive setting 1.2 mV)

Appropriate VF episode detection at 1.2 mV
97.6% (120/123)

VT/VF detection in follow-up

VT/VF episodes appropriately detected and treated in 94 patients; no under-detected episodes
670

Safety

Inappropriate shock · 12 and 24 months

Inappropriate shock rate at 12 months
2.7%
Inappropriate shock rate at 24 months
3.8%

P-wave oversensing (PWOS) adverse events

Patients with AE related to PWOS; none associated with inappropriate shock
4 (0.6%)

T-wave oversensing (TWOS) adverse events

Patients with AE related to TWOS
4 (0.6%)
TWOS with inappropriate shock
3 (0.5%)

Original publication

Sensing and detection performance of the novel, small-diameter OmniaSecure defibrillation lead: in-depth analysis from the LEADR trial. Europace. Published 25 March 2025. doi:10.1093/europace/euaf062

Summarised from the published abstract; every number is checked against it