Arrhythmias & electrophysiologyGuideline or statement

Strategies to improve atrioventricular synchrony in patients with a Micra atrioventricular (AV) leadless pacemaker

EuropacePublished

  • Expert consensus statement

Recommendations

As issued by the authors

  1. Evaluate mechanical atrial activity on transthoracic echocardiography prior to implant

  2. Turn off Auto A3 Threshold and Tracking Check in all patients

  3. Turn off AV conduction mode switch in all patients with high degree AV block

  4. Program the lower rate to 50 bpm, with exceptions based on individual clinical assessment

Original publication

Strategies to improve atrioventricular synchrony in patients with a Micra AV leadless pacemaker. Europace. Published 7 March 2024. doi:10.1093/europace/euae060