Arrhythmias & electrophysiology

Superior vena cava isolation using a pentaspline pulsed-field ablation catheter: feasibility and safety in patients undergoing atrial fibrillation catheter ablation

EuropacePublished Prospective cohort study

Authors' conclusion

SVC isolation using a pentaspline PFA catheter is feasible and safe.

Study context

Population
Consecutive patients undergoing pulsed-field ablation (PFA)-based AF catheter ablation
Participants
105 patients
Outcome
Feasibility (acute SVC isolation) and safety of PFA-based SVC isolation
Methodological detailsconsecutive patients
  • Consecutive patients

Main finding

As reported in the abstract

Acute SVC isolation

Acute superior vena cava (SVC) isolation achieved

105/105

100%

Applications to isolation

6 ± 1

Observational data: associations, not effects.

Safety

Transient phrenic nerve stunning

Transient phrenic nerve stunning
67/105 (64%)
Phrenic nerve palsy at end of procedure and at discharge
  • No phrenic nerve palsy at the end of the procedure and at hospital discharge

Transient high-degree sinus node dysfunction

Transient high-degree sinus node dysfunction
5/105 (4.7%)
  • No recurrence at the end of the procedure and until discharge

Complications · 3-month follow-up

No complication at 3-month follow-up