Valvular & structural heart diseaseRandomised controlled trial

Transcatheter Closure of Patent Foramen Ovale With a Novel Biodegradable Device: A Prospective, Multicenter, Randomized Controlled Clinical Trial

CirculationPublished

  • Randomised controlled trial
  • multicentre
  • non-inferiority

Authors' conclusion

The novel biodegradable PFO closure device, which disappears on echocardiography within 24 months after implantation, demonstrates noninferiority to the traditional nitinol device in both efficacy and safety.

Study context

Population
Patients with patent foramen ovale (PFO)
Participants
190 patients
Comparison
Biodegradable PFO closure device vs nitinol device
  • Biodegradable devicen = 96
  • Nitinol devicen = 94reference
Primary endpoint
PFO closure success rate by contrast echocardiography
Follow-up
6 months postprocedure

Primary result

As reported in the abstract

PFO closure success · 6 months

lower limit of CI −8.98%

Further findings

events

OutcomeBiodegradable deviceNitinol device
Death, embolism, device thrombus or erosion · entire study period00

Device disappearance on transthoracic echocardiography (biodegradable device) · 24 months

Hyperechoic area began decreasing within the first year and disappeared on echocardiography by 24 months

Safety

Surgical device removal (intraprocedural deformation)

Patients in the trial (biodegradable) group requiring surgical device removal because of intraprocedural deformation
1

Original publication

Transcatheter Closure of Patent Foramen Ovale With a Novel Biodegradable Device: A Prospective, Multicenter, Randomized Controlled Clinical Trial. Circulation. Published 13 October 2025. doi:10.1161/circulationaha.125.074609

Summarised from the published abstract; every number is checked against it