Arrhythmias & electrophysiology

Transvenous extraction of conduction system pacing leads: An international ulticenter (TECSPAM) study

Heart RhythmPublished Retrospective multicentre study

Authors' conclusion

The overall success rates of the extraction of CSP leads were very high (although the LBBAP lead dwell time was <3 years), with a low need for extraction tools and minimal complication. Reimplantation in the conduction system is feasible and safe.

Study context

Population
Patients undergoing extraction of conduction system pacing (CSP) leads
Participants
341 patients
Exposure
His bundle pacing vs left bundle branch area pacing (LBBAP)
  • His bundle pacingn = 224
  • LBBAPn = 117
Outcome
Safety, feasibility and clinical outcomes of CSP lead extraction
Methodological details10 international centers
  • 10 international centers

Main finding

As reported in the abstract

Procedural success

Complete procedural success

338

99%

Clinical success

100%

Retained distal fragments

3

1%

Observational data: associations, not effects.

Further findings

Extraction method, lead dwell time >6 months (n = 226)

Manual extraction successful
198 (87%)
Mechanical tools
22 (10%)
Laser
6 (3%)
Femoral tools necessary
3

Reimplantation in the conduction system

Successful reimplantation
233 of 244 attempted (95%)

Safety

Minor complications
7 (2.1%)

Original publication

Transvenous extraction of conduction system pacing leads: An international ulticenter (TECSPAM) study. Heart Rhythm. Published 17 May 2024. doi:10.1016/j.hrthm.2024.04.054

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.