Coronary disease & ACSCohort study

Ultra-Low-Dose Noise-Free Technology to Reduce Radiation Exposure During Coronary Angiography

Circ Cardiovasc IntervPublished

  • comparative study on a single-center, real-world, observational cohort

Authors' conclusion

The Noise-Free technique reduced radiation exposure across cumulative and rate metrics while preserving diagnostic image quality, with the greatest reduction observed in complex cases. These findings support its clinical utility for lowering radiation risk.

Study context

Population
Patients undergoing coronary angiography or percutaneous coronary intervention (PCI) (diagnostic coronary angiography, single-vessel PCI, or chronic total occlusion PCI) using 4 different angiography systemsprocedures in 380 patients · coronary angiography 102 · single-vessel PCI 78 · chronic total occlusion PCI 200
Participants
380 patients
Exposure
ultra-low-dose technique vs the other 3 angiography systems
Outcome
Procedural radiation exposure (dose area product, air kerma, dose rate), stratified by exposure mode and procedure type

Main finding

As reported in the abstract

GroupCoronary angiography total air kermaCoronary angiography total dose area productCoronary angiography fluoroscopy dose rateCoronary angiography cine dose rate
ultra-low-dose Noise-Free technology60.3 (38.5-74.1) mGy4.0 (2.9-5.2) Gy·cm 27.9 (6.0-9.6) mGy/min51.7 (37.4-64.9) mGy/min
Coronary angiography reductions than others: total air kerma
40% to 70%
Coronary angiography reductions than others: total dose area product
37% to 66%
Coronary angiography reductions than others: fluoroscopy dose rate
16% to 62%
Coronary angiography reductions than others: cine dose rate
25% to 79%
Chronic total occlusion PCI reduction relative to the other 3 systems: total air kerma
59% to 73%
Chronic total occlusion PCI reduction relative to the other 3 systems: total dose area product
54% to 69%
Chronic total occlusion PCI reduction relative to the other 3 systems: fluoroscopy dose rate
22% to 64%
Chronic total occlusion PCI reduction relative to the other 3 systems: cine dose rate
60% to 79%

Observational data: associations, not effects.

Secondary outcomes

Baseline patient characteristics, procedural time, operator experience

No significant difference

Image quality (expert semi-quantitative assessment and contrast-to-noise ratio)

Comparable image quality between the 4 systems
  • secondary end point

Original publication

Ultra-Low-Dose Noise-Free Technology to Reduce Radiation Exposure During Coronary Angiography. Circ Cardiovasc Interv. Published 21 April 2026. doi:10.1161/circinterventions.126.016542

Summarised from the published abstract; every number is checked against it