Arrhythmias & electrophysiologyCohort study

Ultrasound-guided parasternal tunneling during intermuscular subcutaneous implantable cardioverter-defibrillator implantation is associated with lower shock impedance and more favorable implant geometry

Heart RhythmPublished

  • retrospective single-center cohort study

Authors' conclusion

FUSIM strategy was associated with lower high-voltage shock impedance and more favorable radiographic parameters. A prospective multicenter evaluation is warranted to determine whether FUSIM reduces the need for routine VF conversion testing.

Study context

Population
Adults undergoing subcutaneous implantable cardioverter-defibrillator implantation (January 1, 2014, to December 2024)Ventricular fibrillation (VF) conversion testing attempted in 214 (80.8%), high-voltage shock impedance available in 186, quantitative radiographic measurements available in 225 · SC 87 · FIM 118 · FUSIM 60
Participants
265 adults
Exposure
FUSIM vs conventional approaches
  • SCn = 87
  • FIMn = 118
  • FUSIMn = 60
Outcome
High-voltage shock impedance during VF conversion testing

Main finding

As reported in the abstract

FUSIM vs SC: P < 0.001 · FUSIM vs FIM: P < 0.01

Observational data: associations, not effects.

Further findings

Impedance ≥70 Ω

First-shock conversion failure at 65 J

3 of 214 (1.4%), all in the SC group
3 of 214 (1.4%)

PRAETORIAN scores and coil-sternum distance

FUSIM group showed lower coil-sternum distance
P < 0.01
Intermuscular (IM) approaches had lower PRAETORIAN scores than SC
  • radiographic measures assessed in a consecutive single-operator subset

Original publication

Ultrasound-guided parasternal tunneling during intermuscular subcutaneous implantable cardioverter-defibrillator implantation is associated with lower shock impedance and more favorable implant geometry. Heart Rhythm. Published 30 March 2026. doi:10.1016/j.hrthm.2026.03.1909

Summarised from the published abstract; every number is checked against it