Unipolar voltage mapping to predict lesion durability during pulsed field ablation

Heart RhythmPublished

  • Patients undergoing pulsed field ablation (PFA) using a pentaspline catheter were enrolled

Authors' conclusion

A unipolar voltage ≤0.4 mV and loss of the terminal S wave on the u-EGM identify regions of durable ablation. Unipolar mapping may provide critical insights into durability and guide more effective PFA.

Study context

Population
Patients undergoing PFA using a pentaspline cathetergroup 1 10 · group 2 41 · group 3 69 · group 4 21
Participants
10 / 41 / 69 / 21
Comparison
Double applications vs single applications
  • group 1n = 10
  • group 2n = 41
  • group 3n = 69
  • group 4n = 21

Main finding

As reported in the abstract

S-wave loss

Unipolar voltage ≤0.4 mV optimally predicted S-wave loss

≤0.4 mV

Further findings

Lesion size · larger and more durable lesions with double applications

P = 0.0075

Temporal stability of core and halo regions

halo regions demonstrated recovery
>1.08 mV unipolar, >0.5 mV bipolar
Core lesions defined by this threshold remained temporally stable

Esophageal S-wave re-emergence

Esophageal S-wave re-emergence occurred in single but not double applications, suggesting reversible electroporation

Conduction recovery in redo procedures

a unipolar voltage ≥0.4 mV independently predicted areas of conduction recovery
≥0.4 mV

Original publication

Unipolar voltage mapping to predict lesion durability during pulsed field ablation. Heart Rhythm. Published 23 January 2026. doi:10.1016/j.hrthm.2026.01.025

Summarised from the published abstract; every number is checked against it