Arrhythmias & electrophysiologyDiagnostic or prognostic study

Use of three-dimensional electroanatomic mapping for epicardial access: needle tracking, electrographic characteristics, and clinical application

EuropacePublished

  • Feasibility and validation study (animal and human)

Authors' conclusion

By tracking NT in the 3D EAM system and continuously monitoring the NT electrogram, it is feasible and safe to obtain EA with minimum or no fluoroscopic guidance.

Study context

Population
Pigs and patients undergoing epicardial access (pericardiocentesis)13 pigs · 22 patients
Participants
13 vs 22
Comparison
3D electroanatomic mapping (EAM)-guided epicardial access with needle-tip electrogram
  • Pigsn = 13
  • Patientsn = 22
Outcome
Feasibility of epicardial access guided by 3D EAM with needle-tip electrogram

Performance

As reported in the abstract

Epicardial access success
  • Epicardial access successfully obtained without complication in 13 pigs and 22 patients

Further findings

Needle tip (NT) potential amplitude, pigs (mediastinum / fibrous pericardium / pericardial sac)

Mediastinum
3.2 ± 1.0 mV
Fibrous pericardium
5.2 ± 1.6 mV
Pericardial sac
9.8 ± 2.8 mV
  • all P ≤ 0.001

NT potential amplitude, patients (mediastinum / fibrous pericardium / pericardial sac)

Mediastinum
1.54 ± 0.40 mV
Fibrous pericardium
3.61 ± 1.08 mV
Pericardial sac
7.15 ± 2.88 mV
  • all P < 0.001

Fluoroscopy time by case order

Animals 1-4 / 5-8 / 9-13
64 ± 15, 23 ± 17, and 0 s; P = 0.01
Patients 1-7 / 8-14 / 15-22
44 ± 23, 31 ± 18, 4±7 s; P < 0.001
Epicardial access (EA) without X-ray exposure
five pigs and seven patients

Safety

events

OutcomePigsPatients
Complications00

Original publication

Use of three-dimensional electroanatomic mapping for epicardial access: needle tracking, electrographic characteristics, and clinical application. Europace. Published 8 April 2024. doi:10.1093/europace/euae089

Summarised from the published abstract; every number is checked against it