Arrhythmias & electrophysiologyDiagnostic or prognostic study

Value of high-output pace-mapping of the right phrenic nerve for enabling safe radiofrequency ablation of atrial fibrillation: insights from three-dimensional computed tomography segmentation

EuropacePublished

  • Diagnostic accuracy study
  • consecutive patients
  • CT segmentation merged with electroanatomical map

Authors' conclusion

Non-capture at 50 mA/2 ms demonstrated very high specificity for predicting a distance to the RPN > 10 mm, ensuring safe radiofrequency delivery.

Study context

Population
Consecutive patients undergoing atrial fibrillation (AF) radiofrequency ablation1,033 pacing sites analysed
Participants
45 patients
Comparison
Right phrenic nerve (RPN) capture threshold on high-output pace-mapping vs CT anatomic RPN location
Outcome
Distance between endocardial capture sites and CT anatomic RPN location, by capture threshold; prediction of distance >10 mm

Performance

Prediction of distance to RPN > 10 mm by RPN-t

AUC

0.846

CI 0.821–0.870

P < 0.001

By group

RPN-t > 20 mA

80.8%

Sensitivity · Specificity 77.5%

RPN-t > 30 mA

68.0%

Sensitivity · Specificity 91.6%

Non-capture at 50 mA

42.4%

Sensitivity · Specificity 97.6%

As reported in the abstract

Distance pacing site to RPN when RPN-t ≤10 mA
7.5 ± 3.0 mm (min 1)
Distance pacing site to RPN with non-capture at 50 mA
19.2 ± 6.5 mm (min 9.4)

Original publication

Value of high-output pace-mapping of the right phrenic nerve for enabling safe radiofrequency ablation of atrial fibrillation: insights from three-dimensional computed tomography segmentation. Europace. Published 31 July 2024. doi:10.1093/europace/euae207

Summarised from the published abstract; every number is checked against it