Arrhythmias & electrophysiology

Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers

J Am Coll CardiolPublished Pooled analysis of prospective clinical trials

Authors' conclusion

The safety and efficacy of chronic LP retrieval of an active helix-fixation device was demonstrated in an international clinical trial setting, with implant durations up to 9 years.

Study context

Population
Patients with an active helix-fixation leadless pacemaker (LEADLESS II phase 1 IDE, LEADLESS Observational, LEADLESS Japan trials) with an attempted leadless pacemaker (LP) retrieval ≥6 weeks post-implantation234 retrieval attempts
Participants
233 patients
Outcome
Retrieval success rate of all attempted chronic LP retrievals
Methodological detailsmulticentre, international
  • Multicentre
  • International

Pooled result

As reported in the abstract

Chronic LP retrieval success

Overall retrieval success rate (205 of 234 attempts)

87.6%

Failed retrievals due to inability to access LP docking button (25 of 29 attempts)

86.2%

Further pooled results

Time from implantation to attempted retrieval

Mean time from implantation to attempted retrieval (range 0.2-9.0 years)
3.2 years
Time to retrieval did not significantly influence retrieval success
P = 0.71

Reason for failed retrieval

Inability to access the LP docking button (25 of 29 failed attempts)
86.2%

Safety

Retrieval-related complications (adjudicated by independent CEC)

11 complications in 9 subjects (9 of 233 subjects)
3.9%

Original publication

Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers. J Am Coll Cardiol. Published 8 January 2025. doi:10.1016/j.jacc.2024.10.094

Summarised from the published abstract; every number is checked against it