Arrhythmias & electrophysiologyMeta-analysis

Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers

J Am Coll CardiolPublished

  • Pooled analysis of prospective clinical trials
  • multicentre
  • international

Authors' conclusion

The safety and efficacy of chronic LP retrieval of an active helix-fixation device was demonstrated in an international clinical trial setting, with implant durations up to 9 years.

Study context

Population
Patients with an active helix-fixation leadless pacemaker (LEADLESS II phase 1 IDE, LEADLESS Observational, LEADLESS Japan trials) with an attempted leadless pacemaker (LP) retrieval ≥6 weeks post-implantation234 retrieval attempts
Participants
233 patients
Outcome
Retrieval success rate of all attempted chronic LP retrievals

Pooled result

As reported in the abstract

Chronic LP retrieval success

Overall retrieval success rate (205 of 234 attempts)

87.6%

Failed retrievals due to inability to access LP docking button (25 of 29 attempts)

86.2%

Further pooled results

Time from implantation to attempted retrieval

Mean time from implantation to attempted retrieval (range 0.2-9.0 years)
3.2 years
Time to retrieval did not significantly influence retrieval success
P = 0.71

Reason for failed retrieval

Inability to access the LP docking button (25 of 29 failed attempts)
86.2%

Safety

Retrieval-related complications (adjudicated by independent CEC)

11 complications in 9 subjects (9 of 233 subjects)
3.9%

Original publication

Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers. J Am Coll Cardiol. Published 8 January 2025. doi:10.1016/j.jacc.2024.10.094

Summarised from the published abstract; every number is checked against it