Arrhythmias & electrophysiology

Continuous Atrial Fibrillation Monitoring From Photoplethysmography: Comparison Between Supervised Deep Learning and Heuristic Signal Processing

JACC Clin ElectrophysiolPublished Diagnostic accuracy study

Authors' conclusion

DNNs perform at least as well as SP models, classify more data, and thus may be better for continuous PPG monitoring.

Study context

Population
Free-living patients with 4 weeks of continuous PPG and ECG monitoring
Participants
204 patients
Comparison
Convolutional deep neural network (DNN) vs signal processing (SP) heuristic for continuous atrial fibrillation (AF) detection from PPG
Outcome
AF detection performance (ROC AUC, sensitivity, specificity) and proportion of data classified
Methodological detailsexternal validation
  • External validation

Performance

As reported in the abstract

AF detection, holdout validation
Test or modelAUCSensitivitySpecificityData classified
SP0.97299.6%94.4%62%
DNN0.97392.295.5%95%

AUC

SP0.972
DNN0.973

Sensitivity

SP99.6%
DNN92.2

Specificity

SP94.4%
DNN95.5%

Data classified

SP62%
DNN95%

Further findings

AF detection, external validation cohort (different population and PPG sensor)

AUC SP 0.989 · DNN 0.994

Data analysed

AF detection, external validation cohort (different population and PPG sensor)

DNN performed better and analysed more signal

Original publication

Continuous Atrial Fibrillation Monitoring From Photoplethysmography: Comparison Between Supervised Deep Learning and Heuristic Signal Processing. JACC Clin Electrophysiol. Published 7 February 2024. doi:10.1016/j.jacep.2024.01.008

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.