Arrhythmias & electrophysiology
Effect of posture on position of subcutaneous implantable cardioverter-defibrillator
Authors' conclusion
A supine or arms-down position during chest radiography can result in lower PRAETORIAN scores and underestimation of associated risk on defibrillation testing failure.
Study context
- Population
- Patients with a subcutaneous ICD
- Participants
- 30 patients
- Exposure
- Lateral chest radiograph standing with arms down or supine vs standard standing with arms elevated
- Outcome
- Agreement in PRAETORIAN score (weighted κ) vs standard position
Main finding
As reported in the abstract
Agreement in PRAETORIAN score vs standard position
| Group | Value | PRAETORIAN score decreased | Changed to lower risk category | High to low risk | PRAETORIAN score ≥90 |
|---|---|---|---|---|---|
| Arms down | κ 0.677 | 10 patients (33%) | 4 | ||
| Supine | κ 0.399 | 16 patients (53%) | 7 | 1 | |
| Standard position | 8 events | 8 of 30 |
Value
Arms downκ 0.677
Supineκ 0.399
Standard position8 events
PRAETORIAN score decreased
Arms down10 patients (33%)
Supine16 patients (53%)
Standard position— not stated in the abstract
Changed to lower risk category
Arms down4
Supine7
Standard position— not stated in the abstract
High to low risk
Arms down— not stated in the abstract
Supine1
Standard position— not stated in the abstract
PRAETORIAN score ≥90
Arms down— not stated in the abstract
Supine— not stated in the abstract
Standard position8 of 30
Observational data: associations, not effects.