Arrhythmias & electrophysiology

Effect of posture on position of subcutaneous implantable cardioverter-defibrillator

Heart RhythmPublished Cross-sectional study

Authors' conclusion

A supine or arms-down position during chest radiography can result in lower PRAETORIAN scores and underestimation of associated risk on defibrillation testing failure.

Study context

Population
Patients with a subcutaneous ICD
Participants
30 patients
Exposure
Lateral chest radiograph standing with arms down or supine vs standard standing with arms elevated
Outcome
Agreement in PRAETORIAN score (weighted κ) vs standard position

Main finding

As reported in the abstract

Agreement in PRAETORIAN score vs standard position
GroupValuePRAETORIAN score decreasedChanged to lower risk categoryHigh to low riskPRAETORIAN score ≥90
Arms downκ 0.67710 patients (33%)4
Supineκ 0.39916 patients (53%)71
Standard position8 events8 of 30

Value

Arms downκ 0.677
Supineκ 0.399
Standard position8 events

PRAETORIAN score decreased

Arms down10 patients (33%)
Supine16 patients (53%)
Standard position— not stated in the abstract

Changed to lower risk category

Arms down4
Supine7
Standard position— not stated in the abstract

High to low risk

Arms down— not stated in the abstract
Supine1
Standard position— not stated in the abstract

PRAETORIAN score ≥90

Arms down— not stated in the abstract
Supine— not stated in the abstract
Standard position8 of 30

Observational data: associations, not effects.

Original publication

Effect of posture on position of subcutaneous implantable cardioverter-defibrillator. Heart Rhythm. Published 3 August 2024. doi:10.1016/j.hrthm.2024.07.122

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.