Arrhythmias & electrophysiology

Electrode Thermal Profile During Ventricular Radiofrequency (RF) Ablation: A Novel Indication of Electrode Tissue Interface to Inform Energy Delivery

JACC Clin ElectrophysiolPublished Mechanistic study

Authors' conclusion

Electrode thermal profiles provide information complementary to force vector and impedance, reflecting electrode-tissue interactions that are also related to discrepancies in impedance fall and temperature during RF ablation.

Study context

Population
Patients undergoing ventricular RF ablation with a catheter with 6 distal electrode temperature sensors2,085 RF applications
Participants
100 patients
Outcome
Relationships between thermal profiles, catheter orientation, impedance fall, steam pops and discordant temperature-impedance changes

Main finding

As reported in the abstract

Thermal profile aligned with force vector at stable sites
60.9%
Peripheral profiles oblique/parallel
72.1%
Central profiles perpendicular
80.4%
Impedance fall discordant with temperature, sites
890 (42.7%)
Catheter orientation and thermal profiles differed for low temperature-high impedance fall vs high temperature-low impedance fall discordant sites
P < 0.001
Steam pops with embedded thermal profile (all sensors ≥45 °C) vs other profiles
8.6% vs 1.3%; P < 0.001

Original publication

Electrode Thermal Profile During Ventricular RF Ablation: A Novel Indication of Electrode Tissue Interface to Inform Energy Delivery. JACC Clin Electrophysiol. Published 24 October 2025. doi:10.1016/j.jacep.2025.09.008

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.