Valvular & structural heart disease
Expanding the Spectrum of Transcatheter Edge-to-edge Repair (TEER) Suitability: Evidence From the EXPAND G4 Post Approval Study
Authors' conclusion
Patients previously deemed TEER unsuitable can be safely and effectively treated with the mitral TEER fourth-generation device.
Study context
- Population
- Patients with mitral regurgitation (MR) treated with the MitraClip G4 System
- Participants
- 1,164 patients
- Comparison
- TEER outcomes across HVC TEER-suitability categories
- Risk of stenosis (RoS)n = 56
- RoIRn = 54
- MMRn = 326
- TEER-suitable (TS)n = 303reference
Methodological detailsglobal, multicentre, postapproval, independent core laboratory
- Global
- Multicentre
- Postapproval
- Independent core laboratory
Primary result
As reported in the abstract
MR reduction (≤1+; ≤2+ for RoIR) · 30 days
Key notes
- RoIR value is MR ≤2+
Further findings
NYHA class I or II, 30 days vs baseline
30 days
Baseline
All-cause mortality · 30 days
| Outcome | RoS | RoIR | MMR | TS |
|---|---|---|---|---|
| Change in KCCQ score · 30 days | +27 ± 26 | +16 ± 26 | +19 ± 26 | +19 ± 24 |
- All-cause mortality — Major adverse events low in all groups
- <3%
Original publication
Expanding the Spectrum of TEER Suitability: Evidence From the EXPAND G4 Post Approval Study. JACC Cardiovasc Interv. Published 26 June 2023. doi:10. 1016/ j. jcin. 2023. 05. 014
Composed by AI from the published abstract and checked against it. How Briefs are made →