Factors affecting electrogram sensing in an insertable cardiac monitor: Insights from surface electrocardiogram mapping analysis

Heart RhythmPublished Mechanistic study

Authors' conclusion

Longer vector length and oblique implant angle yielded the best electrogram sensing and are relevant considerations for ICM implantation procedures.

Study context

Population
Participants undergoing precordial single-lead surface electrocardiogram (ECG) mapping; subset of 50 with clinically indicated insertable cardiac monitor (ICM) implant (Reveal LINQ/BIOMONITOR III, 1:1)1,800 tracings
Participants
150 participants
Comparison
ICM sensing vector length (75 vs 45 mm), vector angle and posture on surface ECG mapping
Outcome
P-wave amplitude (visibility threshold > 0.015 mV)
Methodological detailsblinded analysis
  • Blinded analysis

Main finding

As reported in the abstract

P- and R-wave amplitude and P-wave visibility

P < 0.0001

Median P-wave amplitude larger with 75 mm vs 45 mm vector length
45%
Median R-wave amplitude larger with 75 mm vs 45 mm vector length
53%

Further findings

P-wave amplitude and visibility by BMI category

Longer vector length improved P-wave amplitude and visibility in all BMI categories

Correlation of ICM electrogram vs surface ECG amplitudes

P-wave intraclass correlation coefficient
0.74
R-wave intraclass correlation coefficient
0.80
  • Moderate correlation

Original publication

Factors affecting electrogram sensing in an insertable cardiac monitor: Insights from surface electrocardiogram mapping analysis. Heart Rhythm. Published 13 June 2023. doi:10.1016/j.hrthm.2023.05.034

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.