Arrhythmias & electrophysiology

Human use conditions in fatigue of intravenous leads used in cardiac implanted electronic devices

Heart RhythmPublished prospective, observational, multicenter clinical study

Authors' conclusion

HUCS is the first study to measure alternating in vivo stresses applied to leads from multiple manufacturers that spanned a range of stiffnesses and thereby enable the development of an evidence-based lead-testing standard. The greatest alternating stress occurs in the extravenous region. There, lead stiffness predicts how much of the lead is highly stressed but does not predict maximum alternating curvature.

Study context

Population
Subjects with transvenous leads from 4 manufacturers spanning the range of stiffness in clinical use, imaged by biplane cinefluoroscopy during cardiac motion and in the shoulder region during arm motionsubjects whose images were analyzed
Participants
109 subjects
Exposure
Alternating curvature across lead regions and in relation to lead stiffness
Outcome
Maximum alternating curvature, determinants within each lead region

Main finding

As reported in the abstract

  • Determined primarily by between-patient differences, not differences in lead stiffness

Observational data: associations, not effects.

Further findings

Maximum alternating curvature across regions

Greatest in the extravenous region
P = 0.001

Length of the highly stressed portion of the lead in the extravenous region

Correlated inversely with lead stiffness
P < 0.001

Original publication

Human use conditions in fatigue of intravenous leads used in cardiac implanted electronic devices. Heart Rhythm. Published 3 July 2025. doi:10.1016/j.hrthm.2025.06.050

Composed by AI from the published abstract and checked against it. How Briefs are made →