Arrhythmias & electrophysiology

Impact of Median Sternotomy on Safety and Efficacy of the Subcutaneous Implantable Cardioverter Defibrillator

Circ Arrhythm ElectrophysiolPublished Retrospective cohort study

Authors' conclusion

Sternotomy before or after S-ICD does not confer additional risk relative to a historical control group without sternotomy.

Study context

Population
Patients with S-ICD undergoing sternotomy before or after implantation52±16 years, 47 women
Participants
196 patients
Exposure
Sternotomy before or after S-ICD vs historical nonsternotomy controls
  • S-ICD after sternotomyn = 166
  • Sternotomy after S-ICDn = 30
Outcome
S-ICD lead damage during sternotomy, inappropriate shocks from sternal-wire noise, and failure to terminate spontaneous or induced ventricular arrhythmias
Methodological detailsmulticentre, historical control cohort
  • Multicentre
  • Historical control cohort

Main finding

As reported in the abstract

S-ICD first shocks successfully terminated spontaneous ventricular arrhythmias

31 of 32 patients

97%

Defibrillation threshold testing at implant: first shock success

91%

Inappropriate shocks (most commonly T wave oversensing, n=14)

22 patients

Observational data: associations, not effects.

Safety

Inappropriate shocks · median follow-up 29 months

Patients with inappropriate shocks
22
Related to T wave oversensing
14
  • No difference vs nonsternotomy controls

Original publication

Impact of Median Sternotomy on Safety and Efficacy of the Subcutaneous Implantable Cardioverter Defibrillator. Circ Arrhythm Electrophysiol. Published 24 July 2023. doi:10.1161/circep.123.011867