Arrhythmias & electrophysiologyMechanistic study

Impact of wall thickness on the tissue cooling effect of cryoballoon ablation

EuropacePublished

  • Mechanistic study
  • 3D-printed left atrial-PV model
  • porcine thigh muscle

Authors' conclusion

The cooling effect of cryoballoon freezing is weaker in the deeper layers.

Study context

Population
3D-printed left atrial-PV model (patient data) with porcine thigh muscle, 37°C water tank with pulmonary vein (PV) flow5 freezes per thickness
Comparison
Porcine muscle thickness 2 vs 4 vs 6 mm
Outcome
Tissue cooling effect: average temperature of 12 thermocouples on the side opposite the balloon (roof of left superior PV)
Follow-up
180 s

Main finding

As reported in the abstract

Average tissue temperature

2 mm

−41.4

mean · SD 4.2 · °C

4 mm

−33.0

mean · SD 4.0 · °C

6 mm

8.0

mean · SD 8.7 · °C

P for trend < 0.0001

Further findings

Maximum temperature drop per 5 s (steeper in thin muscle)

Maximum temperature drop per 5 s, 2, 4 and 6 mm thickness in that order
5.2 ± 0.9°C, 3.9 ± 0.7°C, and 1.3 ± 0.7°C
Tissue temperature drops steeper in thin muscle
P for trend < 0.0001
  • P for trend <0.0001

Original publication

Impact of wall thickness on the tissue cooling effect of cryoballoon ablation. Europace. Published 20 May 2024. doi:10.1093/europace/euae135