Arrhythmias & electrophysiology

Improved diagnostic performance of insertable cardiac monitors by an artificial intelligence-based algorithm

EuropacePublished Diagnostic accuracy study

Authors' conclusion

The new AI-based ILR-ECG-A lowered the rate of FP ICM diagnoses significantly while retaining a > 98% sensitivity.

Study context

Population
Recipients of insertable cardiac monitors (ICM) followed by the Implicity monitoring platform
Participants
546 recipients · 879 episodes
Comparison
Artificial intelligence (AI)-based ILR-ECG Analyzer (ILR-ECG-A) reclassification vs ICM diagnoses, against adjudication committee
Outcome
Diagnostic specificity and sensitivity of ILR-ECG-A vs adjudication committee (AC) adjudication
Methodological detailsOne episode per ICM abnormal diagnosis per patient, to avoid clusterization
  • One episode per ICM abnormal diagnosis per patient, to avoid clusterization

Performance

As reported in the abstract

Diagnostic performance of ILR-ECG-A vs AC adjudication
Test or modelEpisodes classified abnormalAdjudicated artefactsAdjudicated false positive (FP)Adjudicated abnormal
ICM87980 (9.1%)283 (32.2%)516 (58.7%)

ICM

Episodes classified abnormal879
Adjudicated artefacts80 (9.1%)
Adjudicated false positive (FP)283 (32.2%)
Adjudicated abnormal516 (58.7%)

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Improved diagnostic performance of insertable cardiac monitors by an artificial intelligence-based algorithm. Europace. Published 3 January 2024. doi:10.1093/europace/euad375

Summarised from the published abstract; every number is checked against it