Arrhythmias & electrophysiology

Inappropriate Therapy and Shock Rates Between the Subcutaneous and Transvenous Implantable Cardiac Defibrillator: A Secondary Analysis of the PRAETORIAN Trial

Circ Arrhythm ElectrophysiolPublished Secondary analysis of the PRAETORIAN randomised trial

Primary result · Inappropriate therapyResult — go to the resultHazard ratio 0.99 95% CI 0.65–1.52

Authors' conclusion

There was no significant difference in IAT and IAS rates between the S-ICD and TV-ICD in a conventional ICD population, but causes and predictors for IAT differed between the devices. After the first IAS, an intervention significantly reduced the recurrence rate of IAS.

Study context

Population
Patients with an indication for implantable cardiac defibrillator (ICD) therapymodified ITT
Participants
849 patients
Comparison
Subcutaneous ICD (S-ICD) vs transvenous ICD (TV-ICD)
  • S-ICDn = 426
  • TV-ICDn = 423reference
Primary endpoint
Inappropriate therapy (IAT)
Follow-up
48 months
Methodological detailsinternational, multicentre, ICD programming mandated by protocol
  • International
  • Multicentre
  • ICD programming mandated by protocol

Primary result

Inappropriate therapy · 48-month Kaplan-Meier cumulative incidence

Hazard ratio

0.99

95% CI 0.65–1.52

P = 0.97

Events by treatment group

S-ICDvs TV-ICD
HR 0.99 (95% CI 0.65–1.52); P = 0.97

As reported in the abstract

Further findings

Independent predictors of inappropriate therapy (IAT) (TV-ICD)

HR (95% CI)P value
Baseline heart rate >80 bpm1.99 (1.05–3.76)P = 0.03
History of AF2.66 (1.41–5.02)P = 0.003
Smoking2.46 (1.31–4.09)P = 0.005

Predictor of IAT by cardiac oversensing (S-ICD)

HR (95% CI)P value
QRS duration >120 ms3.13 (1.34–7.31)P = 0.008
OutcomeS-ICDTV-ICD
IAT episodes, IAS and inappropriate ATP
IAT episodes8189
IAS124130
Inappropriate ATP1124
Main cause of IAT episodes40/8183/89
Causecardiac oversensingsupraventricular tachycardias

Inappropriate shock (patients)

No significant difference between groups
P = 0.14

Inappropriate shock (IAS) recurrence after post-IAS intervention

Post-IAS interventions significantly reduced IAS recurrence in both groups
P = 0.046

Original publication

Inappropriate Therapy and Shock Rates Between the Subcutaneous and Transvenous Implantable Cardiac Defibrillator: A Secondary Analysis of the PRAETORIAN Trial. Circ Arrhythm Electrophysiol. Published 3 December 2024. doi:10.1161/circep.124.012836

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.