Arrhythmias & electrophysiologyRandomised controlled trial
Lattice-tip vs. standard irrigated focal-tip catheter for radiofrequency ablation of the cavotricuspid isthmus - the LINEAR randomized trial
- Randomised controlled trial
- multicentre
- two centres
Authors' conclusion
The lattice tip catheter resulted in higher acute procedural success for radiofrequency CTI ablation compared to the standard irrigated focal-tip catheter.
Study context
- Population
- Patients undergoing radiofrequency cavotricuspid isthmus (CTI) ablation
- Participants
- 102 patients
- Comparison
- Lattice-tip vs Standard
- Primary endpoint
- Achievement and persistence of bidirectional CTI block, confirmed by high-density electroanatomical mapping and adenosine testing
- Follow-up
- After a 60-minute waiting period
Primary result
As reported in the abstract
P = 0.002
Secondary outcomes
First-pass block
P = 0.001
Ablation time for CTI block
P < 0.001
| Outcome | Lattice-tip | Standard | P value |
|---|---|---|---|
| Number of lesions | 8.3 ± 2.4 | 13.4 ± 4.5 | P < 0.001 |
Safety
Procedural complications
- No procedural complications documented
Original publication
Lattice-tip vs. standard irrigated focal-tip catheter for radiofrequency ablation of the cavotricuspid isthmus - the LINEAR randomized trial. Europace. Published 14 April 2026. doi:10.1093/europace/euag046
Summarised from the published abstract; every number is checked against it