Arrhythmias & electrophysiology

Leadless Atrial Pacing Targeting Bachmann Bundle for Atrial Resynchronization: A Single Center Registry

Heart RhythmPublished Single center registry

Authors' conclusion

The BBLAP implant technique presented in this registry is feasible and reduced IACD. Further study on the effect of BBLAP on AF burden is needed.

Study context

Population
Adults with sinus node dysfunction and baseline P wave duration >120ms undergoing Bachmann bundle leadless atrial pacemaker implantationenrolled, 17 had successful BBLAP
Participants
21 adults
Exposure
Baseline and BBLAP left atrial conduction time (LACT) and P wave duration (PWD), axis, and morphology were compared.
  • BBLAPn = 17
Outcome
BBLAP implant success
Methodological detailsprospectively enrolled, consecutive adults
  • Prospectively enrolled
  • Consecutive adults

Main finding

As reported in the abstract

Successful BBLAP

17

81%

Observational data: associations, not effects.

Further findings

PWD, among those with successful BBLAP · reduction with BBLAP

reduction 37 (±13) ms · P < 0.001

LACT · decreased with BBLAP

mean difference 35 (±15) ms · P < 0.001

P wave axis with BBLAP

Inferiorly directed in all cases

Negative terminal P wave in inferior leads at baseline

Corrected with BBLAP
8/8

Coronary sinus (CS) activation shift from proximal-distal to mid- or distal-proximal during BBLAP

Patients with shift
13/17

Safety

No complications in all cases