Arrhythmias & electrophysiology

Left bundle area pacing: Guiding implant depth by ring measurements

Heart RhythmPublished Retrospective cohort study

Authors' conclusion

Ring pacing parameters are more predictive of lead progress than tip measurements. Lead depth and lead-septal angle can be determined from ring impedance measurements.

Study context

Population
Patients with standard pacing indications undergoing left bundle area pacing (LBAP) with SelectSecure 3830 leads via C315 sheaths
Participants
73 patients
Exposure
Successful vs unsuccessful LBAP lead implant attempts
Outcome
Dynamics in pacing parameters (ring and tip measurements)

Main finding

As reported in the abstract

Ring impedance increased stepwise during successful vs unsuccessful attempts
P = .039
Wider lead-septal angle correlated with higher ring impedance
P = .036
Unipolar ring threshold correlated with lead implant depth
P = .029

Observational data: associations, not effects.

Original publication

Left bundle area pacing: Guiding implant depth by ring measurements. Heart Rhythm. Published 21 September 2022. doi:10.1016/j.hrthm.2022.09.013