Arrhythmias & electrophysiology

Left bundle branch pacing lead for sensing ventricular arrhythmias in implantable cardioverter-defibrillator: A pilot study (LBBP-ICD study)

Heart RhythmPublished Pilot study

Authors' conclusion

Sensing from the LBBP lead for arrhythmia detection is safe as ∼90% of the episodes were detected appropriately.

Study context

Population
Cardiac resynchronization therapy (CRT)-eligible patients with successful left bundle branch pacing (LBBP)-optimized implantable cardioverter-defibrillator (ICD) or LBBP-optimized CRT with defibrillator
Participants
30 patients
Outcome
Long-term safety and efficacy of the LBBP lead for appropriate detection of ventricular arrhythmia and delivery of ATP

Main finding

As reported in the abstract

Ventricular arrhythmic events detected · mean follow-up 22.9 ± 12.5 months
Ventricular arrhythmic events documented
254
Appropriate events
n = 225 [89%]
Nonsustained ventricular tachycardia (VT)
n = 212 episodes [94%]
VT
n = 8 [3.5%]
VF
n = 5 [2.5%]
Antitachycardia pacing (ATP) efficacy in terminating VT
75%
Episodes inappropriately detected (T-wave oversensing)
Eleven percent of episodes (n = 29)

Further findings

LVEF

LVEF before LBBP
29.9% ± 4.6%
LVEF after LBBP
43.9% ± 11.2%
Change in LVEF
P < 0.0001

Safety

Inappropriate therapy (ATP)

Episodes with inappropriate ATP
14 episodes (5.5%)

Worsening of tricuspid regurgitation

Patients with worsening tricuspid regurgitation
Three patients (10%)

Original publication

Left bundle branch pacing lead for sensing ventricular arrhythmias in implantable cardioverter-defibrillator: A pilot study (LBBP-ICD study). Heart Rhythm. Published 22 December 2023. doi:10.1016/j.hrthm.2023.12.009

Summarised from the published abstract; every number is checked against it