Long-term outcomes of the pentaspline pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation: results of the prospective, multicentre FARA-Freedom Study

EuropacePublished Prospective non-randomised single-arm study

Authors' conclusion

In this prospective, non-randomized, multicentre study, PVI using a pentaspline PFA catheter was effective in treating PAF patients despite rigourous endpoint definitions and high monitoring compliance and demonstrated favourable safety.

Study context

Population
Patients with paroxysmal AF undergoing de novo pulmonary vein isolation (PVI)
Participants
179 patients
Primary endpoint
Primary safety: composite of device- or procedure-related serious adverse events to 7 days post-ablation and PV stenosis or AE fistula to 12 months; primary effectiveness: composite of acute PVI and chronic success (freedom from atrial tachyarrhythmia >30 s, AAD use or cardioversion after 3-month blanking, or amiodarone or repeat ablation at any time)
Follow-up
12 months
Methodological detailsnon-randomized, multicentre, 13 centres
  • Non-randomized
  • Multicentre
  • 13 centres

Primary result

As reported in the abstract

Freedom from composite primary effectiveness endpoint

Freedom from composite primary effectiveness endpoint

66.6%

Patients with atrial tachyarrhythmia recurrence

41

Recurrence that was recurrent AF

31

Further findings

Acute PVI at index procedure

All PVs successfully isolated with pentaspline PFA catheter

Procedural times (20-min waiting period)

Procedure time (min)
71.9 ± 17.6
Left atrial dwell time (min)
41.0 ± 13.3
Fluoroscopy time (min)
11.5 ± 7.4

Monitoring compliance

Weekly events monitor compliance
88.4
72-h Holter monitor compliance
90.3%

Safety

Composite primary safety endpoint

Composite safety endpoint (one tamponade, one TIA)
two patients (1.1%)

Coronary spasm, PV stenosis, atrioesophageal (AE) fistula

No coronary spasm, PV stenosis or AE fistula

Transient phrenic nerve palsy

Transient phrenic nerve palsy, all resolved during index procedure
four cases

Original publication

Long-term outcomes of the pentaspline pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation: results of the prospective, multicentre FARA-Freedom Study. Europace. Published 22 February 2024. doi:10.1093/europace/euae053

Summarised from the published abstract; every number is checked against it